14. TECHNICAL SPECIFICATIONS
14.1. TECHNICAL FEATURES
Accuracy is indicated as [% of reading + number of digits]. It refers to the following
atmospheric conditions: a temperature of 23°C ± 5°C with a relative humidity < 60%.
14.1.1. Safety Test functions
LOW: 200mA CONTINUITY TEST (AUTO, RT+, RT- MODE)
(*) After Test leads calibration
Test Current > 200mA DC per R5 (Test leads included)
Resolution for Test current: 1mA
Open Circuit Voltage 4V V0 24V
Open circuit Test Voltage <1.3 x Nominal Test Voltage
Short Circuit Current <6.0mA with 500V Test Voltage
Nominal Test Current 500V >2.2mA with 230k
other >1mA with 1k*Vnom
RCD and LOOP function are active only for 50Hz 0,5Hz frequency
PHASE ROTATION : VOLTAGE MEASUREMENT
GROUND TEST: RESISTANCE MEASUREMENT WITH EARTH RODS
Test Current <10mA – 77.5Hz
Open circuit Test Voltage <20V RMS
GROUND TEST: RESISTIVITY MEASUREMENT
20.0 199.9m
200 1999m
2.00 99.99km
100.0 125.6km(*)
0.1m
1 m
0.01 km
0.1 km
(*) setting distance = 10m
Test Current <10mA – 77.5Hz
Open circuit Test Voltage <20V RMS
Comentários a estes Manuais