6.1.5. "AUTO", RT+", "RT-" faulty cases
If the instrument detect
the External Power
supply adapter
connected to instrument
will show the message
displayed to the right.
LOW 05.06.01
-
.- -
R+ R-
--- ---
---mA ---mA
REMOVE POWER
AUTO 0.11
Disconnect the External
Power Supply Adapter
FUNC
CAL
If the terminal voltage is
higher than 15V, the
instrument does not
perform the test and
displays the screen to
the right for 5 seconds.
LOW 05.06.01
-.- -
R+ R-
-.-- -.--
---mA ---mA
VOLT IN INPUT
AUTO 0.11
ATTENTION: the test was
not performed because of
voltage at the terminal
ends.
FUNC
CAL
In the case that:
R
CALIBRATION
>R
MEASURED
the instrument displays
the screen to the right.
LOW 05.06.01
0.00
R+ R-
0.00 0.00
219mA 219mA
CAL > RES
AUTO 0.11
ATTENTION:
R
CALIBRATION
>R
MEASURED
FUNC
CAL