Range Frequency Accuracy
1) 2)
40.00A, 400.0A, 1000A 50 Hz / 60 Hz ± (0.5% rdg + 5d)
40.00A, 400.0A
45 to 50 Hz, 60 to 500 Hz
± (2.0% rdg + 5d)
1000A ± (2.5% rdg + 5d)
40.00A, 400.0A
500 Hz to 3.1 kHz
± (2.5% rdg + 5d)
1000A ± (3.0% rdg + 5d)
ACA AutoVA
TM
Threshold: 1A AC (40Hz ~ 500Hz only) nominal
Crest Factor:
40.00A & 400.0A: < 2.5 : 1 at full scale; < 5.0 : 1 at half scale
1000A: < 1.4 : 1 at full scale; < 2.8 : 1 at half scale
1)
Induced error from adjacent current-carrying conductor: < 0.06A/A
2)
Specified accuracy is from 1% rdg to 100% rdg of range and for measurements made at
the jaw center. When the conductor is not positioned at the jaw center, position errors
introduced are:
Add + 1% rdg to specified accuracy for measurements made WITHIN jaw marking lines
(away from jaw opening)
Add + 4% rdg to specified accuracy for measurements made BEYOND jaw marking lines
(toward jaws opening)
Temperature
Range Accuracy
-50°C to -20°C ± (2.0% rdg + 6°C)
-20°C to 30C ± (2.0% rdg + 3°C)
-5F to -4°F ± (2.0% rdg + 12°F)
-4°F to 572°F ± (2.0% rdg + 6°F)
Type-K thermocouple range & accuracy not included
Frequency
Range 5.00 Hz to 500.0 Hz
Accuracy: ± (0.5% rdg +4d
Range Sensitivity (Sine RMS)
40A > 4A
400A > 40A
1000A > 400A
600V > 30V
THD% rdg-F
1)
(ACD-41PQ only)